State food and drug administration: 15 batches of medical device products failed to qualify
 
Author:中國銘鉉 企劃部  Release Time:2017-8-22 11:34:12  Number Browse:819
 
Medical network on August 22 - in order to strengthen quality supervision and management of medical equipment, security of medical equipment products use is safe and effective, the administration of state food and drug supervision and management of the organization of medical oxygen concentrator (medical oxygen generator), dental non-noble metal alloy casting (including the PFM alloy) and other three varieties 155 (sets) of product quality supervision and checking. The results of the sampling inspection are hereby announced as follows: 
 
1. The sampling inspection project does not meet the standard prescribed medical device products, involving 15 batches of 3 varieties of 9 medical device manufacturers. Specific as follows: 
 
(a) medical oxygen concentrator (medical oxygen generator) 3 enterprises 4 products. 2 medical molecular sieve oxygen producing machines produced by Beijing zhongliancorui technology development co., LTD., the oxygen concentration status indicator is not up to standard; Jiangsu, associating medical technology co., LTD. 1 SD series medical molecular sieve oxygen generator, fuzhou kang medical instrument co., LTD production of 1 medical oxygen concentrator, the influence of back pressure does not conform to the prescribed standards. 
 
(2) 1 batch products of 1 enterprise of dental non-precious metal casting alloy (including ceramic alloy). 1 batch of nickel-base casting alloys produced by Shanghai medical device co., LTD. Dental materials factory. 
 
(3) 10 batches of products from 5 enterprises with sterile vaginal expander. 1 batch of disposable sterile vaginal expander produced by henan yantan industrial co., LTD. 1 batch of disposable sterile vaginal expander produced by hunan ping an medical science & technology co., LTD., the residue of ethylene oxide is not in accordance with the standard; Nanchang city fu kang medical instrument co., LTD., the production of 2 batches the disposable use asepsis vaginal dilator, resistance to deformation ability is not in conformity with the prescribed standards, one batch structural strength is not in conformity with the prescribed standards; Nanchang kang jie medical health products co., LTD., the production of 2 batches the disposable use asepsis vaginal dilator, appearance does not conform to the prescribed standards, one batch resistance to deformation ability is not in conformity with the prescribed standards; Changzhou xiaochun medical equipment co., LTD. Produced 4 batches of disposable sterile vaginal expander, 2 batches of appearance, 2 batches of ethylene oxide residue did not meet the standard requirements. 
 
The above sampling inspection does not meet the specifications of the product. See annex 1. 
 
2. The items that are identified by the sampling inspection project as labels and specifications do not meet the standards specified in the medical device products, and there are 5 varieties of 5 medical device manufacturers, which are specific to: 
 
Medical oxygen concentrator (medical oxygen generator) 5 enterprises 5 products. Hangzhou chong wei air separation technology co., LTD., Beijing quanta jian xiang technology co., LTD., jiangsu jianghang medical equipment co., LTD., one of the production of medical molecular sieve oxygen generator, division (changzhou) electronics co., LTD. Production of 1 small molecular sieve oxygen generator, Shanghai's bright medical equipment co., LTD., production of 1 medical oxygen generator, external markers do not conform to the prescribed standards. 
 
The above sampling inspection does not meet the specifications of the product. See appendix 2. 
 
3. All the items of the sampling inspection project are in conformity with the standard set of medical device products, which involve the three varieties of 54 medical device manufacturers, 135 batches (table), see annex 3. 
 
Four, found in the above sampling does not conform to the standard products, the state food and drug administration has asked enterprises to local food and drug supervision and administration authority in accordance with the regulations on the supervision and administration of medical devices and the food and drug supervision bureau general office notice on further strengthening medical instrument sample work for drug safety (food do machinery supervisor [2016] 9), to investigate the related enterprises. 
 
Relevant dealing with medical device manufacturing enterprise does not conform to the stipulations of the standard products, does not conform to the stipulations of the standard project risk assessment, according to the severity of the medical equipment defect level of recall, is voluntarily recalling and public recall information. The food and drug administration department in the place where the enterprise is located shall supervise the recall of enterprises, and the recalled shall be ordered to be recalled. Provided is not in conformity with the standards of medical equipment products cause harm to human body or there is evidence that may endanger human body health, can take to suspend production, management, use of emergency control measures. The relevant provincial food and drug supervision and administration departments should urge enterprises to find out the reasons as soon as possible, formulate corrective measures and put them in place as scheduled. The relevant disposal situation will be released to the society before September 8, 2017. 
 
Hereby notice. 
 
Annex: 1. The national medical equipment sampling inspection does not meet the standard product list 

序號
標示產品名稱
被抽查單位
標示生產企業
規格型號
生產日期/批號/
出廠編號
抽樣單位
檢驗單位
不符合標準規定項
1
醫用分子篩制氧機 
北京中聯科瑞科技發展有限公司
北京中聯科瑞科技發展有限公司
KR-03A 
2016年5月21日 3A16E0400 
北京市食品藥品監督管理局 
北京市醫療器械檢驗所
氧濃度狀態指示器
2
醫用分子篩制氧機 
陜西省醫療器械公司
北京中聯科瑞科技發展有限公司
KR-03C 
3C16C0563 
陜西省食品藥品監督管理局 
北京市醫療器械檢驗所
氧濃度狀態指示器
3
SD系列醫用分子篩制氧機 
江蘇思締醫療科技有限公司
江蘇思締醫療科技有限公司
SD-05W 
2016.6
52160611210016 
江蘇省食品藥品監督管理局 
江蘇省醫療器械檢驗所
背壓的影響
4
醫用氧氣濃縮器 
福州市瑞康醫療器械有限公司
福州市瑞康醫療器械有限公司
RK-03A 
2016316
RK1600005 
福建省食品藥品監督管理局 
廣東省醫療器械質量監督檢驗所
背壓的影響
5
鎳基鑄造合金 
哈爾濱市德藝齒科技術發展有限公司
上海醫療器械股份有限公司齒科材料廠
1000克/盒 
201506
黑龍江省食品藥品監督管理局 
北京大學口腔醫學院口腔醫療器械檢驗中心
耐腐蝕性
6
一次性使用無菌陰道擴張器 
儋州市第一人民醫院
河南亞都實業有限公司
中號 
616020007 
海南省食品藥品監督管理局 
湖北醫療器械質量監督檢驗中心
抗變形能力
7
一次性使用無菌陰道擴張器 
湖南省腦科醫院
湖南平安醫械科技有限公司
中號 
160310 
湖南省食品藥品監督管理局 
貴州省醫療器械檢測中心
環氧乙烷殘留量
8
一次性使用無菌陰道擴張器 
天津寶坻渤海醫院
南昌市福康醫療器械有限公司
型號:轉軸式;規格:中號 
20151208
天津市市場和質量監督管理委員會 
北京市醫療器械檢驗所
抗變形能力
9
一次性使用無菌陰道擴張器 
定安縣人民醫院
南昌市福康醫療器械有限公司
中號 
2016年01月08日  
海南省食品藥品監督管理局 
上海市醫療器械檢測所
1.抗變形能力;
2.結構強度
10
一次性使用無菌陰道擴張器 
吳忠眾康醫院有限公司
南昌市康潔醫用衛生用品有限公司
型號:A型軸轉式;規格:中號 
20160409 
寧夏回族自治區食品藥品監督管理局 
北京市醫療器械檢驗所
外觀
11
一次性使用無菌陰道擴張器 
隆德閩寧友好醫院
南昌市康潔醫用衛生用品有限公司
A型中號 
20151206 
寧夏回族自治區食品藥品監督管理局 
北京市醫療器械檢驗所
1.外觀;
2.抗變形能力
12
一次性使用無菌陰道擴張器 
馬鞍山市中心醫院
常州曉春醫療器材有限公司
調節式中號 
151201 
安徽省食品藥品監督管理局 
江蘇省醫療器械檢驗所
外觀
13
一次性使用無菌陰道擴張器 
馬鞍山東方婦產醫院
常州曉春醫療器材有限公司
調節式中號(Ⅱ型) 
20150401 
安徽省食品藥品監督管理局 
江蘇省醫療器械檢驗所
外觀
14
一次性使用無菌陰道擴張器 
西北婦女兒童醫院
常州曉春醫療器材有限公司
全透明推拉式中號 
160211 
陜西省食品藥品監督管理局 
江蘇省醫療器械檢驗所
環氧乙烷殘留量
15
一次性使用無菌陰道擴張器 
西北婦女兒童醫院
常州曉春醫療器材有限公司
全透明推拉式中號 
160501 
陜西省食品藥品監督管理局 
江蘇省醫療器械檢驗所
環氧乙烷殘留量
 
 2. National medical device sampling (labeling label specification and other items) does not meet the criteria for product list 

序號
標示產品名稱
被抽查單位
標示生產企業
規格型號
生產日期/出廠編號
抽樣單位
檢驗單位
不符合標準規定項
1
小型醫用分子篩制氧機
杭州創威空分科技有限公司
杭州創威空分科技有限公司
CW0-S5A
20150918
15090309
浙江省食品藥品監督管理局
上海市醫療器械檢測所
外部標記
2
醫用分子篩制氧機
北京廣達健翔科技有限公司
北京廣達健翔科技有限公司
GD-03-C-0-0
2015年11月13日
150605
北京市食品藥品監督管理局
遼寧省醫療器械檢驗檢測院
外部標記
3
醫用分子篩制氧機
成都普龍科學儀器有限公司
江蘇江航醫療設備有限公司
ZH-A36
2016.03.04
1603010251
四川省食品藥品監督管理局
廣東省醫療器械質量監督檢驗所
外部標記
4
小型分子篩制氧機
科邁(常州)電子有限公司
科邁(常州)電子有限公司
FY-3L
2015年12月DE019A1512068
江蘇省食品藥品監督管理局
遼寧省醫療器械檢驗檢測院
外部標記
5
醫用制氧機
安徽紫竹醫療器械有限公司
上海歐亮醫療器械有限公司
OLO-1
20160128
1-1601-352
安徽省食品藥品監督管理局
山東省醫療器械產品質量檢驗中心
外部標記
  
3. National medical device sampling meets the criteria for product list 
 
Food and drug administration 
 
August 14, 2017 

 